Resolution should be chosen via cross-validation

FSC’s standard resolution curve relies on assumptions that real-space masking breaks, so resolution becomes biased. FSCV reframes resolution as hyperparameter selection using two-fold cross-validation over Fourier-shell filters.

Mask-induced shell correlation is the real bias

Masking doesn’t just “hide” parts; it couples Fourier shells, making held-out agreement look stronger than true signal. FSCV addresses this by optimizing a filter against held-out data while only scoring voxels inside the mask.

FSCV recovers trustworthy local resolution and precision

FSCV estimates resolution using an adaptive degrees-of-freedom count tied to mask width, preventing overconfident curves when masks get tight. This keeps sub-region and sharp-mask resolution consistent without relying on noise-substitution diagnostics.