ICP-MS Q&A: Solutions for Mercury, Indium & Gold
Matrix chemistry drives ICP-MS accuracy at trace levels
Correct ICP-MS results depend less on instrument settings than on chemistry: pH, ligands, rinse sequencing, and standard matrix matching control stability, adsorption, precipitation, and carryover.
EDTA can stabilize high-pH metal standards
At basic pH, metals like indium form insoluble hydroxides that distort data even when solids aren’t visible; EDTA’s soluble complexes prevent precipitation but must be matrix-blanked for added background or interferences.
Washout failures can mislead trace conclusions
Mercury persists via volatility-dependent speciation, surface adsorption/desorption, and redox memory effects; rinse effectiveness depends on sequence because strong rinses can leave sulfur/chloride residues or inflate blanks, breaking system suitability and hiding real trace shifts.