Podcast summary

Proficiency Surveys: Testosterone Accuracy and Care Impact

Clinical Chemistry Podcast episode artwork

Clinical Chemistry Podcast

Podcast • 10/28/26 • 13 mins

Listen to the full episode

Traditional PT checks agreement, not clinical accuracy

Commercial guidelines rely on fixed decision thresholds, but nonstandard assay bias makes cutoffs behave differently across methods. Accuracy-based PT uses commutable, reference-calibrated specimens to measure true accuracy; traditional PT mainly shows peer agreement and can miss harmful bias.

Excellent average PT hides method-level threshold shifts

Even with testosterone methods averaging <2% bias versus reference, method groups ranged from −16% to +15%. That implies results near 264 ng/dL could land ~222–304, materially changing diagnosis and treatment based on lab method.

Performance on traditional PT doesn’t reliably predict accuracy

The study tested whether biases on traditional, manipulated PT predict biases on commutable, accuracy-based surveys. Several method groups showed different bias patterns, implying that passing PT acceptance criteria may not reflect accuracy on real patient samples.